FBAC4895-B52C-4E03-B099-A6A8F8B8017B Created with sketchtool. 706D6907-81E1-46A2-B8F9-7B702A1EF2C7 Created with sketchtool. 6A870209-E7F5-44F5-A66F-09E20C85A0DC Created with sketchtool. 0BB04D6C-A9AB-4913-BF4F-5A23954E68E1 Created with sketchtool. 501E26C7-66C1-4DAA-B740-6CFAF55A1B39 Created with sketchtool. C105F7D3-D5EE-43C9-9EE8-05CC42B296D4 Created with sketchtool. D94689D8-7CBB-41C8-AC06-F79D81FF82D5 Created with sketchtool. DA4C7ED9-2512-4D52-94F0-92800F2A6E04 Created with sketchtool. 42E133FD-828F-4CF4-A0BF-560014487A57 Created with sketchtool. F52D9745-4BFD-4506-BEEF-630FD09EF2D9 Created with sketchtool.
Remote stock (5-7 days)
The item is in a remote warehouse, delivery time is 5-7 business days. More info
Delivery options  - Select at checkout
Shipping @price - GLS
12,00
EUR 27,62
Incl. VAT 34,53
Total incl. delivery 49,53
X/Y axis micrometer scales with a scale division of 0.05mm and a 10x10mm reticle

Calibration slide for stereomicroscopes.

The slide has a 10x10mm reticle etched on it. In the center of the slide, there are crosshairs – two micrometer scales of 200 divisions. The scale division is 0.05mm and the length of each scale is 10mm.

The calibration slide (stage micrometer) is used to determine the magnification, linear field of view of the microscope, and the division value of the eyepiece scales and reticles.

With a digital camera, the slide is used to calibrate the image analysis software for each objective.


Calibrating the image analysis software


The calibration is performed for each objective that will be used in observations. The calibration slide is used to determine the calibration factor for the image analysis software.


Calibration procedure:



  1. Fix the calibration slide on the microscope stage.

  2. Select the desired objective and set the maximum camera resolution.

  3. Get a contrast image of the scale on the monitor screen and capture the image.

  4. Select “Calibration” mode in the image analysis software.

  5. Click with the mouse to define the maximum visible distance between the first and last divisions of the scale.

  6. Enter the resulting scale size in real units into the image analysis software.

  7. Enter the calibration name by specifying the current objective magnification.

  8. Check the result: In “Measurement” mode, measure a few divisions of the calibration slide scales in real units.

  9. The image analysis software will save the calibration factor under the selected name. After that, it will automatically convert the results to the selected units of measurement.


Calibration of the eyepiece scale


The calibration of the eyepiece scale is performed for each objective that will be used in observations. The slide should be placed on the stage and the microscope should be focused on the scale. The eyepiece should be rotated in the eyepiece tube until the scales of a slide and an eyepiece are aligned parallel to each other.


Calculation for each objective is carried out according to the formula:

Е=ТL/A, where

E – eyepiece division value

Т – stage division value specified on the stage micrometer (0.05mm)

L – number of stage micrometer divisions

А – number of eyepiece scale divisions.


The resulting values of the eyepiece scale division for each objective shall be recorded.


When measuring an object, first you should count the number of eyepiece scale divisions that the object covers. The result is multiplied by the scale value previously calculated for the selected objective. The resulting value will be the true linear value of the object.

BrandMAGUS
Warranty2 years
EAN5905555017211
Package size (LxWxH)4x3x1 cm
Shipping Weight0.01 kg
Materialoptical glass
Scale length, mm10
Number of scale divisions200
Scale value, mm0.05
Reticle size, mm10х10
Number of reticle cells in a row20
Reticle cell size, mm0.5х0.5
Calibration slide size, mm24