FBAC4895-B52C-4E03-B099-A6A8F8B8017B Created with sketchtool. 706D6907-81E1-46A2-B8F9-7B702A1EF2C7 Created with sketchtool. 6A870209-E7F5-44F5-A66F-09E20C85A0DC Created with sketchtool. 0BB04D6C-A9AB-4913-BF4F-5A23954E68E1 Created with sketchtool. 501E26C7-66C1-4DAA-B740-6CFAF55A1B39 Created with sketchtool. C105F7D3-D5EE-43C9-9EE8-05CC42B296D4 Created with sketchtool. D94689D8-7CBB-41C8-AC06-F79D81FF82D5 Created with sketchtool. DA4C7ED9-2512-4D52-94F0-92800F2A6E04 Created with sketchtool. 42E133FD-828F-4CF4-A0BF-560014487A57 Created with sketchtool. F52D9745-4BFD-4506-BEEF-630FD09EF2D9 Created with sketchtool.
Remote stock (5-7 days)
The item is in a remote warehouse, delivery time is 5-7 business days. More info
Delivery options  - Select at checkout
Shipping @price - GLS
12,00
EUR 59,11
Incl. VAT 73,89
Total incl. delivery 88,89
Magnification: 10x. Field of view: 22 mm. Scale with 0.1mm division

The eyepiece with a scale is used for measuring the size of samples and their structures in linear units: millimeters or microns. The scale is placed in the plane of the field diaphragm with a division of 0.1mm. The eyepiece with a scale is installed instead of a standard 10x eyepiece. A paired eyepiece without a scale is installed in the second eyepiece tube.


The eyepiece is compatible with microscopes that have a 30mm eyepiece tube.


Setting up


Before taking measurements, the eyepiece scale must be calibrated for each objective that will be used. It requires a calibration slide – a stage micrometer with a scale. The minimum division is 0.01mm. The slide should be placed on the stage and the microscope should be focused on the scale. The eyepiece should be rotated in the eyepiece tube until the scales of a slide and an eyepiece are aligned parallel to each other.


Calculation of the eyepiece scale value for each objective


If you use low-magnification objectives, you should calculate how many divisions of the eyepiece scale fit into the slide scale.
If you use medium and high-magnification objectives, you should calculate how many divisions of the slide fit into the scale of the eyepiece.


Calculation for each objective is carried out according to the formula:
Е = ТL/A, where
E – eyepiece division value
Т – stage division value specified on the stage micrometer (0.01mm)
L – number of stage micrometer divisions
A – number of eyepiece divisions.


The resulting values of the eyepiece scale division for each objective shall be recorded.


Calculating the true linear size of an object


When measuring an object, first you should count the number of eyepiece scale divisions that the object covers. The result is multiplied by the scale value previously calculated for the selected objective. The resulting value will be the true linear value of the object.


Compatibility:



  • MAGUS Bio 260T Biological Microscope

  • MAGUS Bio DH260 Biological Microscope

  • MAGUS Bio 270T Biological Microscope

  • MAGUS Bio 290T Biological Microscope

  • MAGUS Bio V360 Biological Inverted Microscope

  • MAGUS Lum 450L Fluorescence Microscope

  • MAGUS Metal V790 DIC Metallurgical Inverted Microscope

  • MAGUS Pol 890 Polarizing Microscope

BrandMAGUS
Warranty2 years
EAN5905555025414
Package size (LxWxH)4x7x4 cm
Lens materialoptical glass
Magnification, x10
Field of view, mm22
Eye relief, mm10
Eyepiece diopter adjustment, diopters±5D
Barrel diameter, mm30
Scale value, mm0.1